7. For every measurement to perform, do the following:
Click on the ”Auto” button. This will evaluate the necessary time to get at list 1000
counts. This value is the minimum to get good statistics on measurements.
Put the probe in the dose rate first listed in the Measures area, as indicated in the
prompt box, then click on “OK” to confirm and run the estimation of the integration
time.
Once the integration time has been estimated, click on the “Start” button. A prompt
box appears, as shown on the following screenshot:
Put the probe in the activity first listed in the Measures area, as indicated in the
prompt box, then click on “OK” to confirm and make the measurement.
While the measurement is being performed, do one of the following:
- Wait until the current measure is complete.
- Click on the “Stop” button to cancel the current measurement.
After measurement, if you have not taken into account the time calculated by the
“Auto” function and set manually the integration time, the following prompt message
may show up to warn about missing statistics.