PTS 400.3 PLUS Operation Manual_R02 Page 239/306
Meter constants of impulse outputs
The following tables are showing the meter constant depending on used current input and internal
voltage range, current range.
Table 2-1: Direct current input 12A
Base constant: cpz
0
= 9000 [imp/Wh], CPZ
0
= 2.5 [imp/Ws]
Range dependent constant of PRS 600.3: cpz = cpz
0
** [imp/Wh]
CPZ = CPZ
0
** [imp/Ws]
Mean frequency at the impulse output: f = CPZ * P, with CPZ = cpz/3600 [imp/Ws]
Maximum frequency (reached at Un, In): fmax = 46.8 kHz
The actual constants CPZx [imp/Ws] and the mean frequency FOutx, x = 1,2,3 of the three impulse
outputs are indicated at the PRS 600.3 in the Reference / setup / impulse output menu.
Note: If a range independent constant C/R is programmed by the user, he must take care, that the
resulting frequency fout stays below 46.8 kHz.
E.g. to test the full range up to 120A, 260V the range independent constant must be:
cpz ≤ 1’800 [imp/Wh] or with unit Wh/imp: cpz 0.0005555 [Wh/imp]
Above annotation is also applying for table 2-2 and 2-3