PTS 400.3 PLUS Operation Manual_R02 Page 264/306
Connections to meter under test (example for IEC-type meter)
Attention! For safety reasons, whenever possible, switch off the mains circuit breaker
during meter manipulations to fix the adapter pins and omega clips for the voltage con-
nections.
Regard the instructions for the use of the installed test terminals and observe local
safety regulations.
Adapter pins
Direct connected
3-wire meter
Installed on site
Maximum test current 120 A
Test links closed
For other type of meters (ANSI Form S, Form A; British Standard BS etc.) consult the documentation
delivered by the manufacturer and adapt the connections to the meter regarding your needs.