Table 5.10—Initial Digit 5, Processor Control Service Codes (PC)
Code Service Code Description
Corrective
Action Code
5002 PC_ERROR_WATCHDOG_SHORT_FAILURE (Main watchdog short test failure. Watchdog failure; param: 0=None, 1=short, 2=long,
3=power.)
1
5003 PC_ERROR_WATCHDOG_LONG_FAILURE (Main watchdog long test failure. Watchdog failure; param: 0=None, 1=short, 2=long,
3=power.)
1
5004 PC_ERROR_WATCHDOG_UNEXPECTED (Unexpected main watchdog reset. Top 16 bits = seconds since last set. Low 16 bits = watchdog
status.)
1
5005 PC_ERROR_RAM_AT_BOOT (RAM error detected during boot; param = Ram Addr of error.)
1
5006 PC_ERROR_BAD_CHECKSUM (Program contents failed Checksum test.)
1
500d PC_ERROR_CONFIG_VERSION (System configuration version mismatch; param = value read.)
1
500e PC_ERROR_CONFIG_CRC (NVRAM system configuration CRC bad; param = value read.)
2, 1
5011 PC_ERROR_METERS_VERSION (System meters version mismatch; param = value read.)
2, 1
5012 PC_ERROR_METERS_CRC (NVRAM system meters/counters CRC bad; param = value read.)
2, 1
5013 PC_ERROR_MFG_DATA_VERSION (Manufacturing data version mismatch.)
2, 1
5014 PC_ERROR_MFG_DATA_CRC (NVRAM manufacturing data CRC bad.)
1
5015 PC_ERROR_FORCED_RESET_FAILED (Forced watchdog reset failed. Unit failed to reset.)
1