LIFEPAK 15 Monitor/Defibrillator
Service Manual
Preventive Maintenance
Cleaning
Section Menu Section Contents Back Index
153
6
Cleaning
Tools and Materials
The tools and materials needed to perform an external cleaning of the LIFEPAK 15 monitor/defibrillator are listed in Table 6.3.
Table 6.3—Cleaning Tools and Materials
Product Description
Static-discharge-protected work area Grounded conductive surface and wrist strap
Isopropyl alcohol
Soap and water
Quaternary ammonium compounds
Peroxide (peracetic acid) solutions
Cotton swabs
Vacuum cleaner
Soft-bristle brush Nonmetallic
Cloth Clean and lint-free
Compressed air Clean and dry (60 psi, max.)