LIFEPAK 15 Monitor/Defibrillator
Service Manual
Replacement Procedures
Saving and Restoring the Setup Configuration
Section Menu Section Contents Procedures Back Index
180
8
Saving and Restoring the Setup Configuration
Before beginning any repair action, the existing setup configuration should be preserved using either of the following methods:
The best method is to store setup options to a computer using the LIFEPAK Defibrillator Software Solutions Configuration Setup
Tool before performing service.
The second method is to print the existing setup configuration, complete repairs, and then manually reconfigure the device.
NOTE: The LIFEPAK Defibrillator Software Solutions Configuration Setup Tool is a Windows®-based application designed to assist
you in managing the setup options in your LIFEPAK 15 monitor/defibrillator. You can download the tool from Physio-
Control.com.
Printing the Setup Configuration
To print the setup configuration:
1. Display the Setup mode (see Setup Mode (p. 92)).
2. Select PRINT DEFAULTS to print the device setup configuration. Save this printout for future reference.
3. Turn the device OFF.
Restoring the Setup Manually
To restore the setup configuration manually:
1. Display the Setup menu.
2. Using the printout from the preceding steps, check the settings in each menu and revise as necessary to match the printout. The
printout items are organized in the same manner as the Setup menu (General, Manual Mode, AED Mode, and so forth). See the
LIFEPAK 15 Monitor/Defibrillator Setup Options instructions provided with your device.
3. Turn the device OFF.