LIFEPAK 15 Monitor/Defibrillator
Service Manual
Battery Maintenance
Storing Batteries
Section Menu Section Contents Back Index
166
7
Storing Batteries
A battery is considered to be in storage when it is not in active use.
LIFEPAK 15 Li-ion batteries require special handling procedures for storage and then placing in use.
Storage temperatures ranges:
~ Short term (<1 week): -20° to 60°C (-4° to 140°F)
~ Long term (>1 week): 20° to 25°C (68° to 77°F)
Operating and storage humidity range: 5% to 95% relative humidity, non-condensing
Fully charge LIFEPAK 15 Li-ion batteries before storing, and at least annually while in storage.
Do not freeze batteries. Damage to the battery may result.
Charge stored batteries before placing in use.
WARNING
POSSIBLE LOSS OF POWER DURING PATIENT CARE Stored batteries lose charge.
Failure to charge a battery before use may cause device power failure without warning.
Always charge a stored battery before placing it in use.