LIFEPAK 15 Monitor/Defibrillator
Service Manual
Troubleshooting
Therapy Service Codes
Section Menu Section Contents Back Index
130
5
Therapy Service Codes
Table 5.13—Initial Character a, Therapy Service Codes (TH)
Code Service Code Description
Corrective
Action Code
a002 TH_ERROR_DEFIB_LINK_DOWN (Lost contact with defib processor. Serial communications link between the main and defib processor is
not functioning.This unit can no longer administer defib therapy. Cycling power may clear the error temporarily, but unit is questionable;
param = last defib message time stamp.)
10
a003 TH_ERROR_PACER_LINK_DOWN (Lost contact with pacer processor. Serial communications link between the main and pacer processor is
not functioning.)
10
a004 TH_ERROR_UNEXPECTED_ENERGY (Unexpected energy in the capacitor. No charge delivered.)
10
a005 TH_ERROR_CAP_OVERCHARGED (Over-charged capacitor.)
10
a008 TH_ERROR_DEFIB_DISABLE (No communication with defib HW.)
10
a00a TH_ERROR_DE_WRONG_ENG_SELECT (Incorrect energy selected; param = energy index.)
7, 10
a00b TH_ERROR_SHOCK_NOT_DELIVERED (Shock not delivered.)
10
a00d TH_ERROR_CHARGING_EXPIRED (Charging time expired.)
10
a00f TH_ERROR_CAP_OUT_OF_RANGE (Capacitor is out of range. Calibration failure.)
10
a010 TH_ERROR_PA_RATE_OUT_OF_RANGE (Pacing rate out of tolerance; param: high-16 = pacer selected range, low-16 = pacer actual rate.)
10
a011 TH_ERROR_PA_CURRENT_OUT_OF_RANGE (Pacing current out of tolerance; param: high-16 = selected current, low-16 is actual
current.)
10