LIFEPAK 15 Monitor/Defibrillator
Service Manual
Replacement Procedures
Warnings and Cautions
Section Menu Section Contents Procedures Back Index
173
8
Warnings and Cautions
The following general warnings and cautions apply to all actions you may perform during maintenance of the LIFEPAK 15 monitor/defibrillator.
DANGER
SHOCK HAZARD Lethal voltages may be present even without operator action. Always
discharge the energy storage capacitor prior to servicing. See the service manual
Capacitor Discharging Procedure (p. 178)” for detailed instructions.
WARNING
SHOCK HAZARD The pacing storage capacitor carry high voltage. Discharge the
capacitors before handling.
POSSIBLE SHOCK AND DEVICE DAMAGE It is possible to pinch and damage wires
during reassembly. To avoid pinching wires, carefully follow reassembly instructions.
CAUTION
POSSIBLE COMPONENT DAMAGE The PCBs contain static-sensitive devices
(SSDs). To avoid damage, observe the special handling practices in Static-Sensitive
Device Handling (p. 174). PCBs contain high impedance circuitry; always handle the
PCB by holding on to the edges.