LIFEPAK 15 Monitor/Defibrillator
Service Manual
Troubleshooting
Printer Service Codes
Section Menu Section Contents Back Index
137
5
Printer Service Codes
Table 5.14—Initial Character b, Printer Service Codes (PR)
Code Service Code Description
Corrective
Action Code
b001 PR_ERROR_TEMP_TOO_LOW (Printhead ADC reading too low; param = ADC value. May be associated with UT_ERROR_DAC_FAILURE and
UT_ERROR_ADC_READ.)
13, 28
b002 PR_ERROR_TEMP_TOO_HIGH (Printhead ADC reading too high; param = ADC value. May be associated with UT_ERROR_DAC_FAILURE
and UT_ERROR_ADC_READ.)
13, 28
b003 PR_ERROR_SELF_TEST_FAIL (Invalid CRC read from HW; param = ADC value.)
13, 28
b004 PR_ERROR_WRONG_PRINTER_TYPE (Saw incorrect printer for config; 50/100 mismatch.)
13, 28