LIFEPAK 15 Monitor/Defibrillator
Service Manual
Troubleshooting
Therapy Service Codes
Section Menu Section Contents Back Index
136
5
a219 PA_ERROR_V_HVIS (HVIS voltage out of range.)
15
a21a PA_ERROR_CAL_CURRENT (Current calibration failed.)
15
a21b PA_ERROR_CAL_Z_300 (Impedance 300 calibration failed.)
15
a21c PA_ERROR_CAL_CURRENT_CRC (Current cal table corrupt.)
15
a21d PA_ERROR_CAL_IMPEDANCE_CRC (Impedance cal values corrupt.)
15
a21e PA_ERROR_CAL_Z_0 (Impedance 0 calibration failed.)
15
a21f PA_ERROR_PACE_I (Current present when not pacing.)
15
a220 PA_ERROR_NO_HVIS_SENSE (No HV present when pacing.)
15
a221 PA_ERROR_EXT_A2D_TEST (External A/D test register reset failed.)
15
a222 PA_ERROR_NO_CAL_HVIS_SENSE (No HV present before current cal.)
15
a223 PA_ERROR_NO_Q4_SENSE (No Q4 sense when not pacing [Q4 shorted].)
15
a224 PA_ERROR_Q4_SENSE (Q4 sense present when pacing [CR25 open].)
15
Table 5.13—Initial Character a, Therapy Service Codes (TH)
Code Service Code Description
Corrective
Action Code