LIFEPAK 15 Monitor/Defibrillator
Service Manual
Preventive Maintenance
Scheduled Replacement Items
Section Menu Section Contents Back Index
149
6
Scheduled Replacement Items
The following items should be replaced at regular intervals because of their finite life span:
• Disposable Electrodes—Replace electrodes by the “Use by” date.
• ECG cable (3-lead/5-wire/12-lead)—Replace to ensure the continued performance of this cable.
• Lithium-ion battery pak — Replace to ensure maximum operating time of the device.
• Coin battery — Replace to ensure the device will not lose battery power for the real-time clock.
NOTE: Contact your Physio-Control Service Representative for assistance when coin battery replacement is required.
Table 6.2 shows the schedule for replacement items.
Table 6.2—Replacement Schedule
Replacement Item Frequency
Replace ECG cable 2 years
Replace battery pak 2 years
Replace coin (clock) battery 5 years