Power Management Service Codes
Table 5.15—Initial Character c, Power Management Service Codes (PM)
Code Service Code Description
Corrective
Action Code
c008 PM_ERROR_LOST_COMMS (Lost communications with power PCB.)
7, 1
c009 PM_ERROR_CLOCK_FAILURE (Power PCB is running on its internal oscillator block because it detected an external clock failure.)
1, 7