Setup Mode
About Setup Mode
Setup mode allows you to change the factory default settings and define custom settings based on local medical protocols and
specific needs. Options include general characteristics, manual and AED mode operating characteristics, alarms setup,
transmission sites, time-of-day clock, and other options. There is also a factory reset option that resets the device to the factory
default settings, (except for transmission menu entries and the maintenance interval, which remain unchanged). When setup is
complete, turn off the device to save the settings. The next time the device is turned on, the operating defaults you selected are
active.
Preserving the Existing Setup Options
Print the existing device setup options by selecting Print Defaults from the Setup mode menu or store setup options using the
LIFEPAK Defibrillator Software Solutions Configuration Setup Tool before performing service.
NOTE: The LIFEPAK Defibrillator Software Solutions Configuration Setup Tool is a Windows
®
-based application designed to assist
you in managing the setup options in your LIFEPAK 15 monitor/defibrillator. You can download the tool from Physio-
Control.com.