Device User Test
Device User test
When you use the SPEED DIAL to navigate to Options/User Test, the device waits until the next self-test cycle is complete. The user
test performs the following tasks:
Performs a self-test
Charges the defibrillation capacitor to 10 joules, then dumps the charge
Stores the test results in the test log
Displays User Test results and prints report
If this operation does not pass, the Service LED illuminates and an error is written to the Service Log (p. 113).