LIFEPAK 15 Monitor/Defibrillator
Service Manual
Preventive Maintenance
Cleaning
Section Menu Section Contents Back Index
154
6
External Cleaning Procedure
Clean the exterior of the device by wiping the surface with any of the following solutions:
Soap and water
Quaternary ammonium compounds
Isopropyl alcohol
Peroxide (peracetic acid) solutions
WARNING
SHOCK OR FIRE HAZARD Do not immerse or soak any portion of this device in water
or any other fluid. Avoid spilling any fluid on the device or accessories.
CAUTION
POSSIBLE CASE DAMAGE Do not clean any part of this device or accessories with
bleach, bleach dilution, or phenolic compounds. Do not use abrasive or flammable
cleaning agents. Do not attempt to sterilize this device or any accessories unless
otherwise specified in the accessory operating instructions.