3. Installation and setup ABL800 FLEX Operator's Manual
Calibration setup, Continued
Calibration
schedule
(continued)
Step Action
3.
Select activity after each measurement by activating the check button
for the desired calibration.
Available
calibration
schedule options
The following options are available.
Option Interval
1 Point calibration 30 min, 1 hour, 2, or 4 hours.
2 Point calibration 1 hour, 2, 4 or 8 hours.
1 Point pH/BG calibration
(for the USA only)
30 min, 1 hour, 2 hours.
tHb calibration Never, 7 days, 1 month, 2, 3, 4 or 6 months
Start time 00:00, 00:15, 00:30, 00:45.……. 23:45. or
12 midnight, 12:15 am, 12:30 am, 12:45
am…….12 midday, 12:15 pm…..11:45 pm.
Activity after measurement none, 1 Point pH/BG calibration (for the USA
only), 1 Point calibration, 2 Point calibration.
Cleaning intervals 8, 24 hours.
3-34