7. Replacements ABL800 FLEX Operator's Manual
Automatic auxiliary programs
Purpose
The following programs might be requested:
Program... Is used...
Rinse To remove traces of a sample, calibrating solution or a quality
control solution from the liquid transport system.
Cleaning To remove lipid deposits from the liquid transport system and
the electrodes by using the Cleaning Solution. The program is
followed by a rinse.
To make the cleaning start automatically – see section
Calibration schedule in Calibration setup in chapter 3 for
details.
Tubing refill To fill the liquid transport system with solution (is performed
automatically during restart after replacement of any solution
container).
Liquid sensor
adjustment
To adjust the liquid sensors in the Wet Section.
Pump
calibration
To calibrate the rotation of the analyzer's pumps (is performed
automatically during restart after replacement of the pump
tubes).
Starting a
program
To start one of the above programs, do the following:
Step Action
1.
Go to the Auxiliary programs menu: Menu - Start Programs -
Auxiliary Programs.
2.
Press the button to start a program.
The following programs can be interrupted by pressing the Stop button:
Interrupting a
program
• Cleaning • Liquid sensor adjustment
• Tubing refill • Pump calibration.
The Rinse program, once started, cannot be interrupted.
7-20

Table of Contents

Other manuals for RADIOMETER ABL800 FLEX

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the RADIOMETER ABL800 FLEX and is the answer not in the manual?

RADIOMETER ABL800 FLEX Specifications

General IconGeneral
BrandRADIOMETER
ModelABL800 FLEX
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals