Note to users of the ABL800 FLEX analyzers
This note to users gathers changes from previous note to users in one document and
outlines some new changes to the operator’s manual of your ABL800 FLEX analyzer
(from software version 6.10).
Please remove the existing note to users from the binder of your manual and place this note
to users in the binder instead.
Changes/Description
Limitations of use and known interfering substances:
CAUTION - Known interfering substances
Substance Interference
ClO
4
–
(drugs)
For ClO
4
–
,
interference on cCa
2+
(1.25 mmol/L
level), cCl
ï€
(110 mmol/L level), and cK
+
(4
mmol/L level)
has been detected:
cCa
2+
(1.25 mmol/L level): ï€0.27*
cCl
ï€
(110 mmol/L level): 4-30
cK
+
(4 mmol/L level): ï€0.3.
Interference –
new interference
results for ClO
4
–
* Depending on the pH level
USB connector
Instead of the CD-ROM drive, a USB connector may be available.
The USB connector can be used for storing data on an USB flash
drive (memory stick) or for connecting USB devices. Can also be
used for installation of software.
Introduction
Instructions to
user
Brief overview
of the change
USB connector or
CD-ROM drive