15. Radiometer settings ABL800 FLEX Operator's Manual
15-18
Calibration verification
Regulations in some countries require verification of the reportable ranges selected
for the measured parameters, and a special feature described below is provided to
assist in the collection of this data.
Once the reportable ranges for the measured parameters are established, the limits
may be entered in the Reportable Ranges program described in Analysis Setup,
chapter 3.
The procedure given below is only a suggestion to easily identify your specific
layout and measuring mode.
Step Action
1.
Make a new report layout called, e.g. "Cal. Verification", based on the
Radiometer default layout – see Patient Report Setup in chapter 3.
2.
Press Patient ID layout and deselect all items except for the Patient
ID and Sample type. Set default sample type to Cal. Verification.
3.
Press Edit Patient Results Layout and deselect all parameters except
for the measured parameters. Deselect range checks for all parameters.
4.
In the Analysis Setup - Syringe Mode Setup dedicate one unoccupied
button to this specific mode (call it, e.g. Cal. Verification) using Edit
name.
Select desired parameters (press Parameters) and layout (press
Layout) that you have made in step 1 above.
Continued on next page
Purpose
Preparatory
steps