2. What Is what ABL800 FLEX Operator's Manual
Analyzer status
Analyzer status
at a glance
The working condition of the analyzer is continuously monitored during its
operation.
To evaluate the analyzer status at a glance before a measurement, use the following
facilities:
parameter bar
color of the traffic light on the Analyzer Status button.
A
nalyzer Status button
Red = cannot be used; Yellow = unreliable; Green = OK
To enter the Analyzer Status screen, press Menu - Analyzer Status.
Analyzer Status
elements
Traffic light color of the Analyzer Status button is determined by the traffic light
colors of the following status elements:
Status element The colors indicate…
Calibration Status
G
REEN – OK.
Y
ELLOW error(s) in the last calibration and/or cal
schedule reminders.
RED – error-prone parameters are repressed.
Continued on next page