3. Installation and setup ABL800 FLEX Operator's Manual
Calibration setup
Programs
Calibration Setup programs are described in this section.
Activate a button to enter the program.
This program allows you to change the default drift tolerances for the pH, pCO
2
,
pO
2
, electrolyte and metabolite electrodes − see Radiometer default settings in
chapter 15 in this manual.
Calibration drift
tolerances
NOTICE: Radiometer recommends the use of the default drift tolerances. Too
narrow drift tolerances will cause electrode drift errors even for normal electrode
fluctuations. If the drift tolerances are made wider, no warning will be given if the
electrode should become unstable. Significant measurement errors could result.
To define your own drift tolerances, do the following:
Step Action
1.
Highlight the desired parameter by touching it on the screen.
The list of available parameters depends on your analyzer version.
2.
Type in your own value and confirm with Enter.
3.
Change other drift tolerances in the same manner.
Press Close to confirm the changes.
4.