EasyManuals Logo

RADIOMETER ABL800 FLEX User Manual

RADIOMETER ABL800 FLEX
438 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #374 background imageLoading...
Page #374 background image
12. Sampling ABL800 FLEX Operator's Manual
12-8
Sampling devices
Radiometer has an assortment of specialized blood gas sampling devices. The
PICO product line is used for puncture and A-line sampling, and CLINITUBES
product line are used for capillary sampling.
For ordering information on sampling devices, see chapter 14: Ordering
Information or contact your local distributor of Radiometer products.
The following arterial blood samplers available from Radiometer are
recommended:

PICO70: Self-filling arterial blood sampler for arterial puncture or A-line
sampling

PICO50 (2 mL): Arterial blood sampler for the aspiration of arterial blood
from A-lines

SafePICO: Self-filling arterial blood sampler for arterial puncture or A-line
sampling (contains a ball for automatic sample mixing in the FLEXQ
sampler tray)
The following CLINITUBES for capillary blood sampling available from
Radiometer are recommended:
 D956G-70-210
 D956G-70-100
 D957P-70-100
 D941G-240-55 (do not use for measuring electrolytes)
 D956G-70-35
(G = glass; P = plastic; the last figure designates volume in L)
CAUTION - Risk of erroneous results
Do not use 85 L CLINITUBES sampling device with high heparin
concentration in FLEXMODE.
Introduction
Arterial blood
samplers
CLINITUBES

Table of Contents

Other manuals for RADIOMETER ABL800 FLEX

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the RADIOMETER ABL800 FLEX and is the answer not in the manual?

RADIOMETER ABL800 FLEX Specifications