7. Replacements ABL800 FLEX Operator's Manual
General information
Viewing
replacement
requirements
To view the replacements that are due, do the following:
Step Action
1.
Check the following on the Analyzer Status screen:
• Replacement schedule for the electrode membranes and the
electrodes
• Messages referring to the membranes/electrodes or user activities
• Traffic light color (G
REEN = No replacements are due at the present
time; Y
ELLOW = A replacement is due)
2.
Highlight the item to enter the screen for replacements of
electrodes/membranes – see the next page.
3.
Press Log Activity to enter the User Activity program to view user
activities list – see chapter 3 in this manual.
4.
Press Close to exit to the Ready screen.
The Hold mode ensures that all wet section activities are suspended and that the
proper restart sequence will follow the exit from the Hold mode.
Entering the
Hold mode
To enter the Hold mode, do one of the following:
• Go to Menu – Analyzer Status – Electrodes and Other. Press Replace.
• Remove the cover to the measuring modules.
• Open the window to the electrode module and wait for draining to complete.
• Remove both inlet flaps.
• Remove a solution container.
Continued on next page