ABL800 FLEX Operator's Manual 3. Installation and setup
Corrective actions
Purpose
This program allows you to do the following:
• to select corrective actions for the events listed in this program,
• to select the traffic light indication, if available, for an event,
• to select analyzer action for the subsequent measurements.
To specify the corrective action and traffic light signal (if available) for each
condition listed in the "Condition" box, do the following:
Step Action
1.
Highlight the desired condition using the up/down arrows in the box.
2.
Select an action for this condition, using the arrow buttons in the
"Corrective action(s)" box – see the table below.
3.
Select the desired traffic light signal (yellow or green, if available) for
the specified event by pressing the traffic light in the "Traffic light
signal" box – see the table below.
4.
Select corrective actions/traffic light signal for the other conditions in
a similar way.
Conditions and
correction
actions
Conditions and corresponding corrective action choices are as follows:
Condition Corrective action Traffic light
Calibration error(s)
present
? on specific parameters
Perform AutoCheck measurement
−
Do not run scheduled AutoCheck
Calibration schedule
reminder(s)
Message on next Patient Result Green or yellow
QC error(s) present ? on specific parameters
−
Continued on next page