3. Installation and setup ABL800 FLEX Operator's Manual
Corrective actions, Continued
Conditions and
corrective
actions
(continued)
Condition Corrective action Traffic light
QC schedule
reminder(s)
Message on next Patient Result Green or yellow
Lock analyzer immediately when
overdue
−
Message on next Patient Result
Green or yellowReplacement
schedule reminder(s)
Lock analyzer when 10 % overdue
−
Message on next Patient Result
Green or yellowSystem message(s)
present
User activity
reminder(s)
Green or yellow
AutoCheck QC
error(s) present
Re-run same level once
−
−
Re-run same level twice
AutoCheck scheduled
level missing
Use ampoule last in schedule Green or yellow
Explanation of
corrective
actions
The explanation of the corrective actions is given below.
Corrective action Explanation
? on specific
parameters
The affected parameter(s) will be marked with “?” in
subsequent patient results.
Message on next
Patient Result
To mark the subsequent patient results on the Message
screen
Lock analyzer
immediately when
overdue
If a quality control measurement is not run 1 minute after
the scheduled time for it, the analyzer will be locked.
As a QC takes more than 1 minute to complete, the
analyzer will enter the "Locked" state during measurement.
It will, however, automatically leave the "Locked" mode
again, if the quality control measurement results are valid.
Lock analyzer
when 10 %
overdue
If a scheduled replacement procedure is more than 10 %
overdue compared with the scheduled time for it, the
analyzer will be locked.
Re-run same level
once
If an error has been registered during a scheduled
AutoCheck measurement, it will be re-run provided the
control solution is available in the carousel.
Continued on next page
3-84