ABL800 FLEX Operator's Manual 13. Specifications
13-13
Analyzer requirements
Power
Rated voltage: 100-240 V; 50/60 Hz; 270 VA.
Maximum voltage fluctuations:  10 %.
Leakage current measured at 230 V and 50 Hz:
Normal condition (ground connected): < 0.005 mA
Single fault condition (ground
disconnected):
< 0.5 mA
Fuses Printer unit has 1 protective fuse: 5 × 20 mm, 4 Amp,
Slow blow (T4AL).
Main fuse includes 2 protective fuses: 5 × 20 mm, 4 Amp,
High break (T4AH). Type Shurter No. 0001.2510.
WARNING – Risk of fire
Fire hazard. Replace fuse as marked.
Start up Cold start: typically 25 minutes
Warm start: 5 minutes
Gas pressure Operating pressure: 1.0-38 bar
Maximum pressure: 38 bar
Location Indoor use only
Relative humidity 20 to 80 %
Operating
temperature
15 ï‚°C to 32 ï‚°C
Storage temperature
–20 C to 60 C
Altitude correction Up to 3000 meters above sea level at standard barometric
pressure of 760 mmHg.
Baro 450-800 mmHg
60.0-106.7 kPa
0.600-1.067 bar
450-800 Torr
Installation category II.
Pollution degree 2.
Analyzer
requirements
Environmental
requirements
Environmental
ratings