ABL800 FLEX Operator's Manual 10. Analyzer shutdown
Standby mode
Purpose
The Standby mode is used for the following:
•
to keep the analyzer ready for measurements (the electrodes conditioned and
solutions thermostatted) during short-term time periods such as weekends or
holidays.
•
to prevent spillage from the waste container when it gets critically full.
By placing the analyzer on standby the consumption of solutions is reduced, as the
scheduled calibrations are suspended.
Entering
Standby
To put the analyzer into Standby mode, do the following:
Step Action
1.
Enter date and time for the analyzer to automatically exit the Standby
mode, using the arrow buttons.
2.
Press Enter Standby to enter the Standby mode. The button changes
to Exit Standby.
Continued on next page