ABL800 FLEX Operator's Manual 2. What Is what
Inlet module
Parts and
functions
The inlet module accepts samples from a syringe/test tube or a capillary (the
syringe and capillary inlet flaps are removed on the picture). Syringe and capillary
inlet flaps are interlocked so that only one may be opened at a time.
Syringe inlet
Syringe inlet flap post
Inlet gasket
unit
Capillary inlet
Inlet probe
Capillary inlet flap post
Preheater
Probe
attachment
Part Function
Probe attachment Secures inlet probe. Open the clip to remove probe.
Preheater
Heats all samples and gases to 37 °C.
Inlet probe Automatically moves into the syringe to aspirate the
required sample.
Capillary inlet flap post Holds the inlet flap for capillary samples.
Syringe inlet flap post Holds the inlet flap for syringe samples.
Capillary inlet Receives sample from a capillary.
Syringe inlet Receives sample from a syringe and test tube.
Inlet gasket unit Provides transfer of samples from a sampling device
to the inlet probe; can be removed.
2-7