ABL800 FLEX Operator's Manual 9. Data management
Calibration Log
Purpose
The Calibration Log is a historical file of calibrations automatically saved in the
Calibration Log after a calibration.
Each calibration report is listed as an individual record. The reports are listed
chronologically, the latest report being at the top of the screen.
Each report is identified as follows (use the left/right scroll bar to see the items):
Item Description
Time Date and time the calibration was performed.
Cal # The ID number of a performed calibration.
Type Indicates the calibration type: 1-point, 2-point or tHb.
Status Indicates the calibration status:
OK
a successful calibration
?
an error detected or a parameter exceeded the
measuring range
Interrupted
a calibration stopped by the operator
Aborted
a calibration stopped by the analyzer most
likely due to missing calibration solution.
NOTICE: The columns displayed in the log and their positions can be changed.
Contact a qualified service technician.
Continued on next page