ABL800 FLEX Operator's Manual 11. Troubleshooting
Leak troubleshooting procedure
Leak error
This condition occurs when a leak detection program has failed, possibly
indicating a leak in the fluid transport system.
To correct the error, do the following:
Procedure
Step Action
1.
Remove the analyzer covers and inspect the fluid transport system for
poor tube connections or worn tubes. Repair as required.
2.
Ensure that electrodes are properly mounted in the measuring chamber
and the electrode connector is latched shut.
3.
Mount all parts and press Restart.
The leak detection is performed automatically during the restart. The
error is corrected when the analyzer enters the Ready mode.
If the error persists, the analyzer remains in the Hold mode. Contact a
service representative.
11-77