ABL800 FLEX Operator's Manual 5. Quality control
AutoCheck measurement
Scheduled
AutoCheck
measurement
A scheduled AutoCheck measurement will start on time provided that no
calibrations are pending. A pending calibration will be performed before the
scheduled AutoCheck measurement.
On the screen that appears when an AutoCheck measurement is about to run
(scheduled either from QC Schedule or from the RADIANCE system), select one
of the three possibilities:
•
No action taken: the AutoCheck measurement starts after 20 seconds (see
counter on the screen)
•
Proceed pressed: the AutoCheck measurement starts immediately.
•
Postpone pressed: the AutoCheck measurement will be delayed by five
minutes.
Unscheduled
AutoCheck
measurement
To start an unscheduled AutoCheck measurement, do the following:
Step Action
1.
Press Analyzer Status – Quality Control.
2.
Highlight the solution to run an unscheduled measurement on.
3.
Press Run AC Ampoule to start the measurement.
The Quality Control Result is shown after the measurement is completed.
NOTICE: It is only possible to choose ampoules that are packed in the carousel. If
an ampoule is not available, the Run AC Ampoule button is inactive.
5-7