5. Quality control ABL800 FLEX Operator's Manual
Manual quality control measurement
Procedure
To make a manual measurement on a QUALICHECK+ or AutoCheck control
solution, do the following:
Step Action
1.
Open the syringe inlet flap.
2.
Place the adapter tip up into the syringe inlet.
3.
Press Ampoule-QC to select the measuring program.
4.
Press Start.
The sample is aspirated and the Quality Control Identification
screen appears. Refer to the next section, for information on entering
data.
5.
When prompted by the analyzer, remove the adapter and close the
syringe inlet flap.
6.
Dispose of the used ampoules as infectious waste*.
* Reference: Clinical laboratory waste management. CLSI document GP5-A2.
High Crea
Check
S8377 Cleaning Met II Solution is used to check the analyzer performance at a
high creatinine level. The check up is made automatically during the system
alignment procedure.
To call an unscheduled High Crea Check, do one of the following:
• Press Menu > Start Programs > Run High Crea Check
• On the Analyzer Status screen highlight the line for Slot 11 and press Run
High Crea Check to start a measurement.
5-6

Table of Contents

Other manuals for RADIOMETER ABL800 FLEX

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the RADIOMETER ABL800 FLEX and is the answer not in the manual?

RADIOMETER ABL800 FLEX Specifications

General IconGeneral
BrandRADIOMETER
ModelABL800 FLEX
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals