ABL800 FLEX Operator's Manual 5. Quality control
Preparing a control solution, Continued
Before measure-
ment on
QUALICHECK+
control solution
(continued)
Step Action
3.
1
2
Place the ampoule in the ampoule opener and break off
the ampoule neck.
4.
Place the ampoule fully into the H700 adapter.
Before manual
measurement on
AutoCheck
control solution
Do the following:
Step Action
1.
Place the ampoule in the H705 Ampoule holder.
2.
Holding the ampoule in its holder between two fingers, shake it
vigorously for at least 15 seconds.
3.
Place the H700 Ampoule adapter over the ampoule and press it down
to open the ampoule.
5-5