ABL800 FLEX Operator's Manual 13. Specifications
13-11
Calibration and maintenance programs
Calibr
ation
Adjustable Duration, min
type frequency
(hours)
ABL837/27/17 ABL835/ 25/15
A
BL830/20/10 ABL805
Cal 1 ½, 1, 2, 4 5
Cal 2 1, 2, 4, 8 10 8
1 Point
pH/BG
(for the
USA
only)
½, 1, 2 2 2 2
tHb Manual -
Duration: 3½ min.
Program Started… Duration, min
ABL837/27/
17
ABL835/25
/15
ABL830/20
/10
ABL805
Cleaning Every 8 or 24
hours, or
called by
operator
4 4¼
Tubing
Refill
After
changing all
solution
bottles or
called by
operator
4¼ 4 4
Decontami-
nation
Called by
operator
4¼ 2¾
Protein
Remover
Called by
operator
3¾ 2½
Liquid
Sensor
Adjustment
Called by
operator
3¼ 3 3 3
Fortsættes på næste side
Calibrations
High Crea
Check
Auxiliary
programs