13. Specifications ABL800 FLEX Operator's Manual
13-10
Sample handling, Continued
Mode Sample
volume
Measuring
time (sec)*
Cycle time
(sec)
Samples
per hour
Syringe – S165
Other fluids – S165
Ampoule - QC
165 L
80
170
21
Capillary – FLEXMODE
Other fluids – FLEXMODE
(no message)
165 L
80 150 24
Capillary – FLEXMODE
Other fluids – FLEXMODE
(message 869 or 870)
30-40 L
or 40-55
L
80 145 25
Capillary – FLEXMODE
Other fluids – FLEXMODE
(message 871 or 872)
50-70 L
80 145 25
Capillary – FLEXMODE
Other fluids – FLEXMODE
(message 873)
65-100
L
80 170 21
Capillary – FLEXMODE
Other fluids – FLEXMODE
(message 874)
> 90 L
100 170 21
Syringe – S95
Capillary – C95
95 L
100
170
19
Syringe – S85
85 L
80 170 21
Capillary – C55
55 L
80 145 25
Capillary – C35 Met
35 L
80 145 25
Syringe – Pleura pH 85 µL 80 170 21
Expired air 15 mL 65 170 21
* From the moment the inlet flap is lifted until the results are displayed.