10. Analyzer shutdown ABL800 FLEX Operator's Manual
Standby mode, Continued
Exiting Standby
The analyzer automatically exits the Standby mode on the selected day and time.
Press Exit Standby to exit standby before the scheduled date/time. If the Exit
Standby button is not shown, you have to logon to the analyzer as a user with
permission to exit standby.
The analyzer will return to the Ready mode upon exit from Standby, unless:
all parameters are marked "Cal Expired"
analyzer is looked due to "QC Expired".
Analyzer is locked due to "Replacement overdue".
Pending calibrations and/or QC measurements will be performed when possible.