ABL800 FLEX Operator's Manual 2. What Is what
Analyzer status, Continued
Analyzer Status
elements
(continued)
Button The colors indicate…
Quality Control
G
REEN OK.
Status
Y
ELLOW error(s) in the last QC measurement and/or
QC schedule reminders.
RED – error-prone parameters are repressed.
Reagents
G
REEN no replacements due at the present time.
Status
Y
ELLOW a replacement is due, Calibration/Cleaning
solutions expire soon or Calibration/Cleaning
solutions have expired.
Electrode and
G
REEN no replacements due at the present time.
others
Y
ELLOW a replacement is due.
System Messages
G
REEN no (critical) messages.
Y
ELLOW – non-critical messages.
R
ED critical messages. The analyzer cannot
calibrate or measure.
AutoCheck
G
REEN no replacements due at the present time.
Status
Y
ELLOW a replacement is due.
Calibrations
status
The Calibrations status gives an overview of the status of the most recently
performed calibration of each type and relevant messages.
The following information is available on the screen:
Calibration Type Lists each calibration type along with its status:
OK
Calibration was accepted.
?
Error(s) detected during calibration.
Pending or overdue calibration. The last
calibration was accepted.
Continued on next page
2-29