3. Installation and setup ABL800 FLEX Operator's Manual
Quality control setup, Continued
QC solutions
(continued)
To assign a control solution to a slot, do the following:
Step Action
1.
Highlight a slot, using the arrow buttons.
2.
For control solutions from Radiometer, scan the bar code or press the
keypad button
to enter the barcode.
3.
For non-Radiometer control solutions, press Add Non-R-.
The warning
Press Accept to continue with the installation of the control solution.
Press Cancel to cancel installation of the control solution.
4.
To delete a control solution, highlight the desired slot and press
Delete. The warning as in step 3 above appears.
Press Accept to continue with the removal of the control solution.
Press Cancel to cancel.
5.
To define whether an AutoCheck quality control measurement is to be
run manually or automatically, press the Manual/Auto button. Useful
if the AutoCheck module is out of order.
The following quality control solutions can be used with the ABL800 FLEX
analyzers:
Quality control system Analyzer
QUALICHECK5+; AutoCheck5+ ABL805
QUALICHECK3+; QUALICHECK5+; AutoCheck3+;
AutoCheck5+
ABL830/20/10
QUALICHECK5+; AutoCheck5+ ABL835/25/15
AutoCheck6+, S8377 Cleaning Met II Solution ABL837/27/17
For the solution types for each quality control system, please refer to chapter 14:
Ordering information in this manual.
You can also use a suitable non-Radiometer quality control system for your
analyzer.
Continued on next page
3-36