EasyManuals Logo

RADIOMETER ABL800 FLEX User Manual

RADIOMETER ABL800 FLEX
438 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #21 background imageLoading...
Page #21 background image
ABL800 FLEX Operator's Manual 1. Introduction
Names and intended use, Continued
Intended use
(continued)
The following parameters can be measured on pleura samples:
Parameter group Parameters
Measured value pH (acidity)
The following parameters can be measured on expired air samples:
Parameter group Parameters
Measured values pCO
2
(carbon dioxide tension)
pO
2
(oxygen tension)
Derived parameters are listed in chapter 13: Specifications and described in detail
in the Reference Manual, chapter 6.
Requirement to
the operator
The analyzer should be used by personnel who have received special education and
training with regard to procedures utilizing in vitro diagnostic medical devices.
Measurements
on animal blood
Animal blood has not been tested on the ABL800 FLEX analyzer. Some
components in animal blood differ from those in human blood, and variations in
the composition of blood from different animal species may also exist.
FLEXMODE
This mode allows you to analyze a blood sample of 35 μL and higher – up to the
maximum volume accepted by your analyzer. Depending on the available sample
volume, the FLEXMODE provides the highest number of parameters: from all
available to as many as reliably possible.
This mode is not available in the ABL8x7 FLEX analyzers.
Measurements
on pleural fluids
Pleura pH can be measured on pleural fluids. Corrections are present in the device.
Other fluids
mode
All parameters available on your ABL800 FLEX analyzers can be measured on
fluids other than heparinized human whole blood
NOTICE: Before using this mode you must establish "user-defined corrections"
for each parameter used, on the fluid in question. The corrections assume a linear
correlation between the measured value and the reference instrument. The data
used for establishing "user-defined corrections" have to cover the desired
measuring range. If no user-defined corrections are entered, you will measure in
this mode as if on heparinized human whole blood.
FLEXQ module
The FLEXQ module can accommodate up to three samplers simultaneously. It
reads a sampler's barcode, mixes the sample and transports the sampler to the inlet
for aspiration and analysis without any further assistance from the operator.
Results can be delivered via FLEXLINK (for information refer to the RADIANCE
Installation and Setup Manual).
NOTICE:
The model ABL810 can also be ordered without oximetry parameters: as ABL810
BG only.
1-3

Table of Contents

Other manuals for RADIOMETER ABL800 FLEX

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the RADIOMETER ABL800 FLEX and is the answer not in the manual?

RADIOMETER ABL800 FLEX Specifications

General IconGeneral
BrandRADIOMETER
ModelABL800 FLEX
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals