ABL800 FLEX Operator's Manual 7. Replacements
Replacing inlet gasket unit and inlet probe, Continued
Replacing the
inlet probe
(continued)
Step Action
5.
Disconnect the tube from the inlet probe and the nipple on the inlet
module.
6.
Connect the new tube to the nipple on the inlet module and to the new
probe.
Then insert the inlet probe with the tube into the groove of the
attachment and seat it properly. Close the clip.
7.
Mount the inlet gasket unit and inlet flaps, then press Restart.
7-13