7. Replacements ABL800 FLEX Operator's Manual
Decontamination and Protein Removal programs, Continued
Protein Removal
program
(continued)
Step Action
3.
Open the syringe inlet flap and place the syringe tip in the inlet.
Press Start to start aspiration of the solution and the program to begin.
4.
5.
Remove the syringe and close the inlet flap as prompted on the screen.
When the program is completed, the analyzer returns to the Ready
screen.
6.
NOTICE: The program can be terminated at any time by pressing the STOP
button.