EasyManua.ls Logo

RADIOMETER ABL800 FLEX - Page 361

RADIOMETER ABL800 FLEX
438 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
ABL800 FLEX Operator's Manual 11. Troubleshooting
Electrode troubleshooting procedures, Continued
To troubleshoot the cGlu and cLac electrodes, do the following:
cGlu and cLac
troubleshooting
procedure
Step Action
1.
Remove the electrode and check the measuring chamber for clots;
clean the measuring chamber with a cotton stick moistened with
distilled water. Make sure that no cotton fibers are left in the
measuring chamber.
2.
Replace the electrode membrane.
3.
Call a 1-point calibration. If the error persists, replace the electrode.
To troubleshoot the cCrea-A and cCrea-B electrodes, do the following:
cCrea electrodes
troubleshooting
procedure
Step Action
1.
Remove the electrodes and check the measuring chamber for clots;
clean the measuring chamber with a cotton stick moistened with
distilled water. Make sure that no cotton fibers are left in the
measuring chamber.
2.
Replace the membranes of both electrodes.
3.
Call a 2-point calibration. If the error persists, replace the electrodes.
Refer to chapter 7 for specific instructions on replacing electrodes and membranes.
Remember to record all replacements.
NOTICE:
11-79

Table of Contents

Other manuals for RADIOMETER ABL800 FLEX

Related product manuals