ABL800 FLEX Operator's Manual 13. Specifications
13-3
Measured parameters, Continued
Parameter Unit Measuring range Test range
FHbF % 0.0-100.0* 0-80
fraction 0.00-1.000* 0.0-0.8
cK
+
mmol/L; meq/L 0.5-25.0 2-8
cNa
+
mmol/L; meq/L 7-350 120-180
cCa
2+
mmol/L 0.20-9.99 0.51-2.2
meq/L 0.40-19.98 1.0-4.4
mg/dL 0.80-40.04 2.0-8.8
cCl
mmol/L; meq/L 7-350 95-150
cGlu mmol/L 0.0-60 0.5-15
mg/dL 0-1081 9.0-270
cLac mmol/L; meq/L 0.0-30 0.5-15
mg/dL 0-270 4.5-135
cCrea mol/L 10-1800 50-1500
mg/dL 0.11-20.4 0.57-17.0
ctBil mol/L 1-1000* 0-400
mg/dL 0.0-58.5* 0.0-23.4
mg/L 0-585* 0-234
* The values are for the analyzer with the activated "Out of range suppression"
function. In case this function has not been activated, the measuring ranges will be
as follows:
Parameter Unit Measuring range
sO
2
, FO
2
Hb, % 2.0 % to 102.0 %
FCOHb, FMetHb fraction 0.020 to 1.020
FHbF % 15 % to 102 %
fraction 0.15 to 1.02
ctBil mol/L 20 to 1000
mg/dL 1.2 to 58.5
mg/L 12 to 585
Continued on next page