ABL800 FLEX Operator's Manual 13. Specifications
13-15
Analyzer specifications, Continued
Number of ampoules in the carousel 0 to 20
Positioning of ampoules in the carousel Random
Lot change 2 Lots of the same level at the
same time possible
Ampoule volume, liquid 0.7 mL
Expiration of ampoules
24 months at 25 ï‚°C (including 15
days at up to 32 ï‚°C)
Conditioning time (from room temperature) < 15 min. with filled carousel
Scanning time < 60 sec with filled carousel
Manual quality control measurement
possible
Yes
Remote control Remote monitoring and start of
measurement via the RADIANCE
system.
The specifications for the FLEXQ module are as follows:
Number of samplers in the sampler tray up to 3
Sampler type safePICO with safeTIPCAP
Positioning of samplers in the sampler tray Random
Sample mixing time 7 sec
Scanning time per sample Typically <1 sec
Warning signal for missing sampler ID < 5 sec
Barcode scanner reads Code 128 between 5-10 mil.
AutoCheck
module
specifications
FLEXQ module
specifications