[SENSe<Ch>:]CORRection:ADVanced:HARMonic:ORDer:ADD <Order>
Adds a harmonic order to the list of harmonic orders at which source and receiver cali-
brations shall be performed for harmonics measurements in channel <Ch>.
Use [SENSe<Ch>:]CORRection:ADVanced:HARMonic:ORDer:REMove to get this
list, and [SENSe<Ch>:]CORRection:ADVanced:HARMonic:ORDer:REMove to
remove a harmonic order from this list.
Use SOURce<Ch>:POWer<PhyPt>:CORRection:HARMonic[:ACQuire] and
[SENSe<Ch>:]CORRection:POWer<PhyPt>:HARMonic:ACQuire) to calibrate.
Suffix:
<Ch>
.
Channel number
Setting parameters:
<Order> Order number
Usage: Setting only
Options: R&S ZNA-K4
Manual operation: See "Harmonic Orders to Calibrate" on page 536
[SENSe<Ch>:]CORRection:ADVanced:HARMonic:ORDer:REMove
Removes a harmonic order from the list of harmonic orders at which source and
receiver calibrations shall be performed for harmonics measurements in channel <Ch>.
Use [SENSe<Ch>:]CORRection:ADVanced:HARMonic:ORDer:REMove to get this
list, and [SENSe<Ch>:]CORRection:ADVanced:HARMonic:ORDer:ADD to add a
harmonic order to it.
Use SOURce<Ch>:POWer<PhyPt>:CORRection:HARMonic[:ACQuire] and
[SENSe<Ch>:]CORRection:POWer<PhyPt>:HARMonic:ACQuire) to calibrate.
Suffix:
<Ch>
.
Channel number
Usage: Event
Options: R&S ZNA-K4
Manual operation: See "Harmonic Orders to Calibrate" on page 536