Command reference
R&S
®
ZNA
1410User Manual 1178.6462.02 ─ 20
Example:
*RST; SOUR:CMOD:PORT3 ON
Reset the analyzer and define the source signal of port 3 as
coherent.
SOUR:CMOD:PORT3:PHAS 90
Assign 90° phase to the voltage of the coherent signal at port 3.
SOUR:CMOD ON
Turn source coherence on.
Options: R&S ZNA-K6
Manual operation: See "Def'd Phase Coherence" on page 648
SOURce<Ch>:CMODe:PORT<Pt>:PHASe
<Value>
Defines the phase (or phase start) for the coherent signal at port no. <PhyPt>.
Use SOURce<Ch>:CMODe:PORT<Pt>:PHASe:SPAN to set the phase span.
Suffix:
<Ch>
.
Channel number. If unspecified the numeric suffix is set to 1.
<Pt> Physical port number
Parameters:
<Value> Port-specific phase value
Increment: 1-deg
*RST: 0 deg
Example: See SOURce<Ch>:CMODe:PORT<Pt>[:STATe] on page 1409
Options: R&S ZNA-K6
Manual operation: See "Phase Start" on page 649
SOURce<Ch>:CMODe:PORT<Pt>:PHASe:SPAN <Value>
Sets the phase span for port <Pt> in channel <Ch>, if source coherence is enabled
Use amplitude imbalance sweep SOURce<Ch>:CMODe:PORT<Pt>:PHASe to set the
start phase.
Does not apply if an imbalance sweep is running.
Suffix:
<Ch>
.
Channel number
<Pt> Port number
Parameters:
<Value> Phase span in degrees
Options: R&S ZNA-K6
Manual operation: See "Phase Span" on page 649
SCPI command reference