EasyManua.ls Logo

R&S ZNA67 - Page 1435

R&S ZNA67
1709 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Command reference
R&S
®
ZNA
1435User Manual 1178.6462.02 ─ 20
SOURce<Ch>:POWer:ALC:STOLerance <Tolerance>
Defines the variation of the ALC-controlled source signal level (Settling Tolerance in
ALC Config dialog) in channel <Ch>.
Suffix:
<Ch>
.
Channel number
Parameters:
<Tolerance> Settling tolerance [dB]
Range: 0.02 to 1
Increment: 0.01
*RST: 0.1
Manual operation: See "Settling Tolerance" on page 481
SOURce<Ch>:POWer<Pt>:ALC:TPOWer <Value>
Defines the target power value the ALC for port <Pt> in channel <Ch> stabilizes at the
selected measurement receiver (SOURce<Ch>:POWer<Pt>:ALC:RECeiver).
Only applies if the target power isn't taken from the port configuration (SOURce<Ch>:
POWer<Pt>:ALC:TPOWer:DEFault on page 1435 OFF).
Suffix:
<Ch>
.
Channel number
<Pt> Port number
Parameters:
<Value> Default unit: dBm
Manual operation: See "ALC: Target Power/Target Power from Port Config"
on page 480
SOURce<Ch>:POWer<Pt>:ALC:TPOWer:DEFault <Boolean>
Defines whether the ALC target power is taken from the port configuration or can be
set using SOURce<Ch>:POWer<Pt>:ALC:TPOWer.
Suffix:
<Ch>
.
Channel number
<Pt> Port number
Parameters:
<Boolean> *RST: ON (1)
Manual operation: See "ALC: Target Power/Target Power from Port Config"
on page 480
SOURce<Ch>:POWer<Pt>:ALC:VNETwork <Usage>
Allows you to use the configured offset or single-ended deembedding network for ALC.