GUI reference
R&S
®
ZNA
481User Manual 1178.6462.02 ─ 20
Remote command:
SOURce<Ch>:POWer<Pt>:ALC:COUPle
Apply Wave Correction
See "Wave Correction" on page 316.
Remote command:
[SENSe<Ch>:]CORRection:EWAVe[:STATe]
Settling Tolerance
Defines the maximum variation of the source signal level after the ALC has settled.
The "Settling Tolerance" has an impact on the ALC Path IF Bandwidth in "Auto" mode:
other things being equal, the smaller the "Settling Tolerance" the smaller the auto-
selected filter bandwidth.
Remote command:
SOURce<Ch>:POWer:ALC:STOLerance
Show Advanced Settings
Shows or hides advanced ALC settings:
ALC Optimize ← Show Advanced Settings
Enables/configures ALC optimization.
"Auto" (default)
The firmware decides whether speed optimization shall be performed
"On"
Optimization active
"Off"
Optimization inactive
Remote command:
SOURce<Ch>:POWer<Pt>:ALC:OPTimize
ALC Path IF Bandwidth ← Show Advanced Settings
Selects the bandwidth in the ALC control loop.
Values above 100 kHz are only available on the wideband signal path. The 7 MHz filter
was introduced in firmware version 2.40 to improve convergence speed, which allows
shorter pulses with ALC.
"Auto" means, the analyzer selects the appropriate IF bandwidth, depending on the
source frequency and level.
Stimulus softtool