GUI reference
R&S
®
ZNA
518User Manual 1178.6462.02 ─ 20
calibration. You can also switch to another channel and configure/start this
channel's calibration (see Chapter 4.2.2.3, "Multi-channel setup dialog",
on page 112).
"Calibrate All" section
The main purpose of the buttons in this section is to perform complex multi-channel
calibrations in an efficient way.
"Quick Start Calibrate All..." immediately starts the calibration, using a previ-
ously configured subset of the available calibration setups, or all available cali-
bration setups.
Opens the calibration wizard that guides you through the calibration data acqui-
sition.
"Configure/Start Calibrate All..." opens the Calibrate All dialog that allows you
to configure calibration setups for an arbitrary set of channels and to enable/
disable a subset of the configured calibration setups. Finally you can acquire
calibration data for the enabled calibration setups in an efficient way.
"Power Cal Only" section
The "Scalar Power Cal" button opens the Power Cal wizard, which allows you to
perform "Ref. Receiver", "Meas. Receiver" and scalar source power ("Source Flat-
ness") calibrations for the current channel.
Guided Calibration...
Use "Start... (Guided)" to open the Guided Calibration wizard that allows you to config-
ure and run a set of calibrations efficiently.
Cal Validation ...
Opens the Cal Validation dock widget that allows you to set up and run a Calibration
validation.
5.11.1.3 Calibration wizard
Guides you through the acquisition of calibration data (system error correction and
power correction).
Access:
Channel – [Cal] > "Start Cal" > "Cal Active Channel" – "Quick Start Calibration"
Calibration setup dialog > "Start Cal"
Channel – [Cal] > "Start Cal" > "Calibrate All" – "Quick Start Calibrate All"
Calibrate All dialog > "Start Cal"
If accessed via "Quick Start Calibration" or the calibration setup dialog, a single chan-
nel is calibrated. If accessed via "Quick Start Calibrate All" or the "Calibrate All" dialog,
possibly multiple channels are calibrated. Depending on the calibration setup, the cali-
bration can also involve several power calibration steps.